Cell Counting Calculator (Hemocytometer)
Fill in the fields and press Calculate to see the result.
About the Cell Counting Calculator (Hemocytometer)
Work out the cell concentration from a hemocytometer count: cells per mL, viability with trypan blue, total cells and the volume to seed a number of cells.
Count the cells in the four large corner squares of a Neubauer chamber and enter the dilution with trypan blue.
How it is calculated
Cells/mL = cells counted ÷ squares × dilution × 10⁴ · Viability = live ÷ (live + dead)
Each large square of a Neubauer hemocytometer is 1 mm × 1 mm with a depth of 0.1 mm, so it holds 0.1 µL, or 10⁻⁴ mL. The average count per square times 10,000 is the number of cells per mL in the chamber, and times the dilution factor the number in the original suspension: mixing cells 1:1 with trypan blue doubles it.
Trypan blue stains dead cells, so the share of unstained cells is the viability. Count 100 to 300 cells for a reliable result, include the cells touching two sides of each square but not the other two, and dilute the sample if the squares are crowded.
Example
220 live and 20 dead cells in 4 squares with a 1:1 trypan blue dilution give 1.1 × 10⁶ live cells per mL and 91.7% viability. To seed 200,000 cells, take 181.82 µL.
Frequently asked questions
How do I calculate cells per mL from a hemocytometer?
Divide the cells counted by the number of large squares, multiply by the dilution factor and by 10,000. 220 cells in 4 squares diluted 1:1: 55 × 2 × 10⁴ = 1.1 × 10⁶ cells/mL.
How do I calculate cell viability?
Divide the live (unstained) cells by all the cells counted and multiply by 100: 220 live and 20 blue cells give 91.7%.